Filtering the protein
To reduce the number of crystals and increase their size, try filtering the protein solution prior to setting up the experiment. Try the following filter sizes: 0.22, 0.1 micron and 300 kD molecular weight cut-off. Try the Millipore centrifugal filters.
Naomi Chayen

Pseudo-Microseeding
When working with crystals that grow fairly rapidly (one day) try the following. Pipette multiple protein drops (2 to 4 works best) onto the cover slide. Using a single pipette tip, get the reservoir solution to mix with the first drop. Now, go back to into the reservoir with the same tip to get the reservoir solution for the second drop. Continue for the remaining drops with the same tip. In certain cases, seeding starts very quickly, so by using the same tip one can introduce minute seeds to successive drops. Use the same cover slide with multiple drops to minimize evaporation.
Mike Sintchak

Low molecular weight PEGs
When screening with low molecular weight PEGs try microbatch. Crystals appear rapidly with PEG 400-2000. To convert to vapor diffusion use 0.2 M buffer in the well and a 1:1 drop ratio. Try using a positive displacement pipette such as the Anachem Microman 1 - 10 microliter. These are much more accurate.
Lesley Haire

Purest is not the best
A protein which was purified and showed some faint bands of contaminants on a native gel was crystallized and solved successfully. The same protein, purified by HPLC and resulting in a single band native gel did not crystallize.
Michal Harel

Cryoprotectant
When making up cryoprotectant solutions containing glycerol, put a test tube of glycerol in a beaker of warm water. The viscosity falls and it is easier to pipette accurately.
Elspeth Garman

Mass spectrometry
We found mass spectrometry like ESMS and MALDI highly efficient in determining impurities and/or microheterogeneities in our protein sample/batch. In most cases it is a simple, straight forward method which requires a minimum amount of sample. In some cases it has shown to detect impurities/microheterogeneities when other techniques did not.
David Leys
TECHNICAL SUPPORT
Tel: 949-425-1321
Fax: 949-425-1611
Email: tech@hrmail.com
LOCATION
Hampton Research
34 Journey
Aliso Viejo, CA 92656-3317
Toll Free: 800-452-3899
Tel: 949-425-1321
Fax: 949-425-1611
BUSINESS HOURS
7:00 am to 5:00pm
Monday - Friday
(Pacific Standard Time)
info@hrmail.com   |  800.452.3899   |  949.425.1321   |  Fax 949.425.1611   |   ©COPYRIGHT 2003 - 2013 HAMPTON RESEARCH CORP. ALL RIGHTS RESERVED.   |  PRIVACY POLICY